Primary service
CMM Calibration and Repair
Request a quotation for performance verification or repair
Request a quotation

Calibration and verification are two different things
Calibration establishes the relationship between a machine's indications and a reference value, with a stated uncertainty. What most shops order is performance verification: a pass/fail test against ISO 10360-2 or ASME B89.4.1, usually paired with geometric adjustment.
Where the controller allows it, an update to the volumetric error-compensation map follows adjustment. Some controllers are OEM-locked and the map cannot be touched from outside.
What a visit covers
Each calibration or reverification visit follows a fixed sequence: scales, geometry, length error, probing error, then as-left verification.
- 01
Scales and reader heads
X, Y and Z scales are inspected and cleaned; reader heads are checked and adjusted for proper output.
- 02
Air bearings and drives
Air bearings, airlines and drive condition are checked before geometry is tested.
- 03
Geometry
Straightness, linearity and squareness are measured in the XY, ZX and ZY planes.
- 04
Length measurement error
E0 and E150 tested per ISO 10360-2: five lengths, seven positions, three repeats, 105 measurements.
- 05
Probing error
PFTU tested on a dedicated calibrated test sphere per ISO 10360-5, separate from the probe's own qualification sphere.
Sequence of work
- 01
Soak
Thermal soak and environment check before any test begins.
- 02
As-found
Length, geometry and probing error recorded before adjustment.
- 03
Adjustment
Geometric correction and error-map update where the controller allows it.
- 04
As-left
Verification repeated and certificate issued with before-and-after data.

Why Cleveland CMM Calibration and Repair
- 01
As-found and as-left both reported
Both data sets appear on the certificate wherever adjustment occurs, supporting a reverse-traceability review.
- 02
Artifacts matched to the test
Gauge blocks, step gauges and ball bars are selected to the length and orientation the standard specifies.
- 03
Repair alongside calibration
Probe heads, scales, reader heads, air bearings, drives and counterbalance systems are serviced on the same visit where practical.
Coverage across Northeast Ohio and the Great Lakes region
On-site scheduling and travel terms vary by distance from Cleveland; confirm coverage for your location before booking.
Other services
Discuss CMM Calibration and Repair Requirements
Describe the machine and the standard it was built to